Publication

Back to overview

Triple echo steady-state (TESS) relaxometry

Type of publication Peer-reviewed
Publikationsform Original article (peer-reviewed)
Publication date 2013
Author Heule Rahel, Ganter Carl, Bieri Oliver,
Project Development of fast quantitative diffusion and transverse relaxation time magnetic resonance steady state imaging concepts for living tissues
Show all

Original article (peer-reviewed)

Journal Magnetic Resonance in Medicine
Volume (Issue) 71(1)
Page(s) 230 - 237
Title of proceedings Magnetic Resonance in Medicine
DOI 10.1002/mrm.24659

Abstract

Purpose Rapid imaging techniques have attracted increased interest for relaxometry, but none are perfect: they are prone to static (B0) and transmit (B1) field heterogeneities, and commonly biased by T 2/T1. The purpose of this study is the development of a rapid T1 and T2 relaxometry method that is completely (T2) or partly (T1) bias-free. Methods A new method is introduced to simultaneously quantify T1 and T2 within one single scan based on a triple echo steady-state (TESS) approach in combination with an iterative golden section search. TESS relaxometry is optimized and evaluated from simulations, in vitro studies, and in vivo experiments. Results It is found that relaxometry with TESS is not biased by T2/T 1, insensitive to B0 heterogeneities, and, surprisingly, that TESS-T2 is not affected by B1 field errors. Consequently, excellent correspondence between TESS and reference spin echo data is observed for T2 in vitro at 1.5 T and in vivo at 3 T. Conclusion TESS offers rapid T1 and T2 quantification within one single scan, and in particular B1-insensitive T2 estimation. As a result, the new proposed method is of high interest for fast and reliable high-resolution T2 mapping, especially of the musculoskeletal system at high to ultra-high fields. © 2013 Wiley Periodicals, Inc.
-